LabGard ES FH NU-565 Class II Type B2 Biosafety Cabinet

The NU-565 sets the bar for Class II Type B2 performance by complying with not only the biological safety cabinet standards of NSF/ANSI 49 but also the fume hood requirements of UL 1805 and ASHRAE 110.

Overview

The NU-565, a Class II Type B2 biosafety cabinet, excels in biological and chemical vapor protection and is tested to NSF/ANSI 49, UL 1805, and ASHRAE 110 standards. Enhanced safety features include a UL 586 fire retardant exhaust HEPA filter, externalized outlets and service valves, and dual-interlocked exhaust flow via a mechanical switch and velocity pressure flow grid. Continuous monitoring through a touch screen ensures operator awareness of the exhaust grid and work area airflow, mitigating risks of exhaust failure.

UL Classified

Experience unparalleled safety with the NU-565, uniquely 'UL Classified' to meet rigorous standards. Its compliance with UL 61010-1 ensures electrical and mechanical reliability, while adherence to UL 1805 guarantees material flame resistance and optimal airflow efficacy, offering users a product and peace of mind.

 


UL 586

Experience unparalleled safety with a UL 586 certified exhaust filter, removing 99.97% of particles 3 microns or larger. This high-efficiency unit doubles as a critical fire line, withstanding up to 700°F (371°C) and halting fire spread through ventilation systems, ensuring your lab's resilience against contaminants and combustion.
UL 586 Biosafety Cabinet Exhaust Filter

Advanced Exhaust Interlock

Secure your lab's safety with an advanced exhaust interlock system, engineered to activate the internal supply blower only with adequate exhaust flow, ensuring proper air barrier inflow velocity from the get-go. This intelligent feature prevents contaminated work zone air from compromising your laboratory environment, reinforcing user protection.

 


Specifications

NU-565 Class II, Type B2 Biosafety Fume Hood
Airflow Control Options Air Supply On/Off (Requires NU-952-012), Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels N/A
Cord Pass Through Standard Right
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) N/A
Outlet(s) Standard One Left Front Faring, Optional One Right Front Farring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Two Valves on Right Side Wall are Standard, Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) Two Couplings on Left Side Wall are Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) N/A
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light N/A
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded (UL 586)
Lighting LED
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 1/2 HP DC ECM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 3-inch (76mm) 99.99% Efficiency on 0.3 microns
Viewing Window Inches (mm) 18.5 (470) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel with Prop-Up Rod
Work Zone Wrap Single Piece Stainless Steel Side and Back Wall Construction
Data Output Features RS-232, USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm)
Est. Shipping Height Inches (mm)
Est. Shipping Width Inches (mm)
Exterior Depth with Armrest Removed Inches (mm) 32 3/8 (823)
Exterior Height (Minimum for Transport) Inches (mm) 62 (1575)
Exterior Height Inches (mm) 62 (1575)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 88 1/2 (2248)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 94 1/2 (2401)
Exterior Width Inches (mm) 53 5/8 (1362)
Nominal Width Feet (Meters) 4 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 20 7/8 (530)
Useable Work Tray Area Width Inches (mm) 45 7/8 (1165)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 26 (660)
Work Area Height Inches (mm) 25 3/16 (640)
Work Area Width Inches (mm) 46 3/8 (1178)
Blower/Lights Amps 2.1
Electrical Configuration 115V60Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 8
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 815 (1385)
Other Built-in Exhaust Option(s) N/A
Plant Duct Static Pressure Inches (mm) 1.7 w.g. (43mm w.g.)
Performance Standard NSF-ANSI-49
Safety Certification UL, UL-C
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 60 fpm (0.30 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 474
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 540 (245)
Net Weight lbs (kg) 490 (222)
Airflow Control Options Air Supply On/Off (Requires NU-952-012), Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels N/A
Cord Pass Through Standard Right
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) N/A
Outlet(s) Standard One Left Front Faring, Optional One Right Front Farring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Two Valves on Right Side Wall are Standard, Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) Two Couplings on Left Side Wall are Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) N/A
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light N/A
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded (UL 586)
Lighting LED
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 1/2 HP DC ECM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 3-inch (76mm) 99.99% Efficiency on 0.3 microns
Viewing Window Inches (mm) 18.5 (470) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel with Prop-Up Rod
Work Zone Wrap Single Piece Stainless Steel Side and Back Wall Construction
Data Output Features RS-232, USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm)
Est. Shipping Height Inches (mm)
Est. Shipping Width Inches (mm)
Exterior Depth with Armrest Removed Inches (mm) 32 3/8 (823)
Exterior Height (Minimum for Transport) Inches (mm) 62 (1575)
Exterior Height Inches (mm) 62 (1575)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 88 1/2 (2248)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 94 1/2 (2401)
Exterior Width Inches (mm) 77 5/8 (1972)
Nominal Width Feet (Meters) 6 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 20 7/8 (530)
Useable Work Tray Area Width Inches (mm) 69 7/8 (1775)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 26 (660)
Work Area Height Inches (mm) 25 3/16 (640)
Work Area Width Inches (mm) 70 3/8 (1788)
Blower/Lights Amps 2.6
Electrical Configuration 115V60Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 10
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 1275 (2166)
Other Built-in Exhaust Option(s) N/A
Plant Duct Static Pressure Inches (mm) 1.8 w.g. (46mm w.g.)
Performance Standard NSF-ANSI-49
Safety Certification UL, UL-C
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 60 fpm (0.30 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 584
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 730 (331)
Net Weight lbs (kg) 680 (308)

Review

Options

  • 8" (203 mm) Access Opening Tempered Sliding Glass, NSF Approved Cord Pass Through Right Side
  • Air Supply On/Off Control
  • Cabinet Off Override
  • Delay Off Exhaust Airflow
  • Exhaust Airflow Loss Override
  • Plastic Armrest
  • Stainless Steel Armrest
  • Bag-In/Bag-Out Exhaust Hepa Filter
  • Standard Option without Bag-In/Bag-Out Exhaust
  • Metal Frame Exhaust Filter UL-586
  • Bottom Right Plugged Coupling
  • IV Bar with 6 Hooks
  • Left Sidewall Cord Pass-Through
  • Stainless Steel Window Edge Protector
  • Worktray Prop-Up
  • Faring Outlets - 115V Left Side With 115V Right Side
  • Faring Outlets - 115V Left Side With Right Side Plate
  • Faring Outlets - GFI Left Side With Right Side Plate
  • Faring Outlets with GFI Right Side and 115V Left Side
  • 1/2" RIGID COPPER
  • 3/8" Black Pipe
  • 3/8" Soft Copper
  • 115V 60Hz 15 Amp Power Input
  • Junction Box
  • Bottom Left Remote Control Valve
  • Bottom Right Remote Control Valve
  • Middle Left Remote Control Valve
  • Middle Right Remote Control Valve
  • Top Left Remote Control Valve
  • Top Right Remote Control Valve
  • Seismic Brackets for Adjustable Basestand
  • Side Mount Seismic Brackets For Adjustable Basestand
  • Side Mount Seismic Brackets For Telescoping Basestand
  • Bottom Left Plugged Coupling
  • Middle Left Plugged Coupling
  • Middle Right Plugged Coupling
  • Side Panels without Plumbing
  • Top Left Plugged Coupling
  • Top Right Plugged Coupling
  • Side Panels with Plumbing Out Back
  • Side Panels with Plumbing Out Bottom
  • Side Panels with Plumbing Out Top
  • 3-inch (76 mm) Metal Frame Supply Separatorless HEPA Filter 99.99% efficient @ 0.3 µ with Permanent Plenum

Request Your Quote for a NU-565

Please fill out the form below for a lab equipment quote and we'll get to work on getting you what you need. Need to talk with someone right away? Give us a call at 763-553-1270.

Thank you for completing the form.