In-VitroCell NU-5800 Direct Heat CO2 Incubator

The In-VitroCell NU-5800 provides the basic functionality to promote tissue cell growth by controlling Temperature and CO2 gas levels in the growth chamber. Relative humidity up to 95% is produced through convention by a water pan. The Closed Loop HEPA Filtration System creates clean conditions to minimize potential contamination entering the growth chamber.

Overview

The NU-5800 CO2 incubator offers touch panel control and monitoring of temperature and CO2 to meet the demands of a broad range of cell culture applications. Beyond using an infrared sensor for precise control of CO2 and separately adjustable door and perimeter heaters to balance internal temperature. The air pump outside the 200L growth chamber also constantly draws air through a HEPA filter and cycles it back to create ISO Class 5 quality air at positive pressure and minimize contamination. A touch screen makes it easy to enter setpoints and view or download historical performance. In addition, this moderately sized incubator can be stacked to efficiently use the valuable space in your laboratory.

Modern User Experience

Experience the simplicity and efficiency with NuAire's NuTouch control system. This intuitive touchscreen interface revolutionizes how you interact with your CO2 incubator. Adjust settings, view historical data, and calibrate with simple taps and on-screen instructions. NuTouch's user-friendly controls ensure maintaining an optimal growth environment for tissue cultures is straightforward and stress-free.

Optimal Temperature Control

The direct heat technology in our incubator ensures a uniform temperature environment, which is crucial for optimal tissue culture growth. With strategically placed heating elements around the chamber, including the door perimeter and the door itself, this design achieves balanced heat distribution within a robust, steel-insulated chamber. The independent adjustment of door and perimeter heaters caters to uniformity, providing a consistent temperature vital for sensitive tissue culture work, even in varying ambient conditions. This precise temperature control ensures an ideal environment for successful tissue culture growth.

Precision IR Sensor: Optimal Growth, Less Calibration

Our innovative IR sensor technology elevates mammalian cell culture growth with unmatched precision. By controlling CO2 levels within the growth chamber to maintain optimal pH, and using wavelengths absorbed only by CO2, the measurements remain unaffected by other components such as water vapor. This advanced design provides stable output, minimizes drift, and reduces the need for frequent calibrations, ensuring consistent, reliable performance for your cell cultures.

Advanced HEPA Filtration

The In-VitroCell's closed-loop HEPA filtration system ensures an optimal mammalian cell culture growth environment. This system continuously samples and filters chamber air through an external HEPA filter, creating a clean air environment with slight positive pressure inside the chamber, similar to an ISO Class 5 clean room. This process significantly reduces contamination risks and minimizes cell desiccation. In-VitroCell offers superior Constant Contamination Control, providing a multi-tiered defense for the safety and integrity of your cell cultures.

Specifications

In-VitroCell NU-5800 CO2 Incubator
CO2 Accuracy ± 0.1%
CO2 Control Logic Fixed Algorithm/Manual, Environmental Adaptable
CO2 Display Resolution 0.1%
CO2 Range 0.1 to 20%, 5% Default Set Point
CO2 Recovery Up to 5% ± 0.2% / -0.5 in 5 minutes Average
CO2 Sensor Type Infrared (NDIR)
Segmented Doors Optional Set of 6 Inner Doors, 2 Across by 3 Down
Access Port Location Right Side
Chamber Air Filter 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay
Exterior Door Seal Magnetic Outer Door Gasket
Exterior Material Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish
In-line Gas Filters 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor
Inner Doors One
Insulation High Density Insulation with High "R" Rating
Interior Door Seal Silicone Rubber Gasket
Interior Material Polished Type 304L 16 Gauge Stainless Steel
Sample Port Location Inner Glass Door
Shelf Capacity lbs (kg) 25 lbs per Shelf / 125 lbs per Incubator
Shelf Material Stainless Steel or Copper Alloy
Shelves 4 Shelves Supplied, 20 Shelves Maximum Capacity
Data Output Features RS-485, USB, 4-20mA
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 31 (787)
Est. Shipping Height Inches (mm) 51 (1219)
Est. Shipping Width Inches (mm) 31 (787)
Exterior Depth Inches (mm) 27.188 (691)
Exterior Height Inches (mm) 39.688 (1008)
Exterior Width Inches (mm) 26.750 (680)
Footprint Depth inches (mm) 20.75 (527)
Footprint Width inches (mm) 23.75 (603)
Interior Depth Inches (mm) 20.053 (509)
Interior Height Inches (mm) 28.500 (724)
Interior Volume Cubic Feet (Liters) 7.06 (200)
Interior Width Inches (mm) 21.257 (540)
Shelf Depth Inches (mm) 18 3/4 (476)
Shelf Width Inches (mm) 19 3/4 (502)
Water Pan Depth 1.5 (38)
Water Pan Length Inches (mm) 12 (305)
Water Pan Width 10 (254)
Electrical Configuration 115V60Hz
Power Cord Length Foot (Meter) 8 (2.5)
Safety Certification UL, UL-C
O2 Accuracy
O2 Control Logic
O2 Display Resolution
O2 Measuring Range
O2 Recovery
O2 Sensor Type
O2 Set Point Range
Decontamination Cycle Not Available
RH Accuracy
RH Control Logic
RH Control Method Water Pan, Convection
RH Control Range
RH Display Resolution
RH Evaporator Box Capacity
RH Recovery
RH Sensor Type
RH Water Pan Capacity
Chamber Air Circulation Method Air Pump
Chamber Jacket Technology Air Jacket via One or Two Axial Fans Drawing Air into Shell
Control System NuTouch
Chamber Heater(s) Location Walls, Top and Bottom of Chamber
Door and Perimeter Heater Control Logic Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable]
Temperature Accuracy °C ± 0.1%
Temperature Control Direct Heat
Temperature Control Range 5°C above ambient (to a 30°C max ambient) to 55°C
Temperature Display Resolution 0.1°C
Temperature Recovery 0.12°C/min. on Average
Temperature Sensor Type Precision Integrated Circuit
Temperature Set-Point Range 5°C to 55°C (37.0°C Default)
Temperature Uniformity °C ± 0.35°C @ 37°C
Est. Shipping Weight lbs (kg) 325 (147)
Net Weight lbs (kg) 235 (106)
CO2 Accuracy ± 0.1%
CO2 Control Logic Fixed Algorithm/Manual, Environmental Adaptable
CO2 Display Resolution 0.1%
CO2 Range 0.1 to 20%, 5% Default Set Point
CO2 Recovery Up to 5% ± 0.2% / -0.5 in 5 minutes Average
CO2 Sensor Type Infrared (NDIR)
Segmented Doors Optional Set of 6 Inner Doors, 2 Across by 3 Down
Access Port Location Right Side
Chamber Air Filter 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay
Exterior Door Seal Magnetic Outer Door Gasket
Exterior Material Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish
In-line Gas Filters 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor
Inner Doors One
Insulation High Density Insulation with High "R" Rating
Interior Door Seal Silicone Rubber Gasket
Interior Material Polished Type 304L 16 Gauge Stainless Steel
Sample Port Location Inner Glass Door
Shelf Capacity lbs (kg) 25 lbs per Shelf / 125 lbs per Incubator
Shelf Material Stainless Steel or Copper Alloy
Shelves 4 Shelves Supplied, 20 Shelves Maximum Capacity
Data Output Features RS-485, USB, 4-20mA
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 31 (787)
Est. Shipping Height Inches (mm) 51 (1219)
Est. Shipping Width Inches (mm) 31 (787)
Exterior Depth Inches (mm) 27.188 (691)
Exterior Height Inches (mm) 39.688 (1008)
Exterior Width Inches (mm) 26.750 (680)
Footprint Depth inches (mm) 20.75 (527)
Footprint Width inches (mm) 23.75 (603)
Interior Depth Inches (mm) 20.053 (509)
Interior Height Inches (mm) 28.500 (724)
Interior Volume Cubic Feet (Liters) 7.06 (200)
Interior Width Inches (mm) 21.257 (540)
Shelf Depth Inches (mm) 18 3/4 (476)
Shelf Width Inches (mm) 19 3/4 (502)
Water Pan Depth 1.5 (38)
Water Pan Length Inches (mm) 12 (305)
Water Pan Width 10 (254)
Electrical Configuration 230V5060Hz
Power Cord Length Foot (Meter) 8 (2.5)
Safety Certification UL, UL-C
O2 Accuracy
O2 Control Logic
O2 Display Resolution
O2 Measuring Range
O2 Recovery
O2 Sensor Type
O2 Set Point Range
Decontamination Cycle Not Available
RH Accuracy
RH Control Logic
RH Control Method Water Pan, Convection
RH Control Range
RH Display Resolution
RH Evaporator Box Capacity
RH Recovery
RH Sensor Type
RH Water Pan Capacity
Chamber Air Circulation Method Air Pump
Chamber Jacket Technology Air Jacket via One or Two Axial Fans Drawing Air into Shell
Control System NuTouch
Chamber Heater(s) Location Walls, Top and Bottom of Chamber
Door and Perimeter Heater Control Logic Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable]
Temperature Accuracy °C ± 0.1%
Temperature Control Direct Heat
Temperature Control Range 5°C above ambient (to a 30°C max ambient) to 55°C
Temperature Display Resolution 0.1°C
Temperature Recovery 0.12°C/min. on Average
Temperature Sensor Type Precision Integrated Circuit
Temperature Set-Point Range 5°C to 55°C (37.0°C Default)
Temperature Uniformity °C ± 0.35°C @ 37°C
Est. Shipping Weight lbs (kg) 325 (147)
Net Weight lbs (kg) 235 (106)

Review

Options

  • Copper Alloy Chamber 115V
  • Copper Alloy Chamber 230V
  • Stainless Steel Chamber 115V
  • Stainless Steel Chamber 230V
  • Inner Glass Door
  • Sectioned Inner Door
  • Left Side Door Hinge
  • Euro Power Cord Option Shipping Kit
  • UK Power Cord Option

Request Your Quote for a NU-5800

Please fill out the form below for a lab equipment quote and we'll get to work on getting you what you need. Need to talk with someone right away? Give us a call at 763-553-1270.

Thank you for completing the form.

We will be in touch. If you have any urgent questions, please reach out at 763-553-1270.