TouchLink Control System
The TouchLink control system calculates inflow velocity based on the data received from the exhaust airflow sensor and displays it alongside the readings from the downflow sensor on the front panel. Touch screen control, historical monitoring, and configurable settings enable users to manage airflow when performance must be demonstrated and recorded.
Thermistor Exhaust Sensor
A single dual-thermistor airflow sensor in the exhaust measures airflow to enable the FlowGard to calculate inflow velocity for users desiring the most straightforward measurement of cabinet containment performance.
Downflow Air Sensor
The downflow sensor measures velocity inside the work zone and passes the data to the TouchLink display to assure users of product protection, a critical aspect in many pharmaceutical and biotechnology applications.
Performance History
The TouchLink control panel showcases a week's worth of cabinet performance data on-screen, while also facilitating the download of a full month's operational history to a USB. This feature is essential for procedures requiring documented airflow conditions in compliance with Good Manufacturing Practice (GMP) standards.
Specifications
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Plate |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.99% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1/2 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 46 (1168) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 87.375 (2219) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 93.375 (2371) |
Exterior Width Inches (mm) | 41 5/8 (1057) |
Nominal Width Feet (Meters) | 3 ft. (0.9m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 33 7/8 (860) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 34 3/8 (873) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 2.3 |
Blower/Lights Amps [8-inch (203 mm) Access Opening] | 2.1 |
Electrical Configuration | 115V 60Hz |
Outlet Amps | 3 |
Power Cord | 14 GA - 3 Wire, 15A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 10 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 55dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 320 (544) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 276-501 (649-851) |
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 269 (457) |
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 255-450 (383-765) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 903 |
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) | 826 |
Heat Rejected Vented (BTU/Hour) | 120 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 460 lbs. / 209 kg. |
Net Weight lbs (kg) | 410 lbs. / 186 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Glass |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.995% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1/2 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.995% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 46 (1168) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 88 3/8 (2245) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 94 3/8 (2397) |
Exterior Width Inches (mm) | 41 5/8 (1057) |
Nominal Width Feet (Meters) | 3 ft. (0.9m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 33 7/8 (860) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 34 3/8 (873) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 1 |
Electrical Configuration | 230V 50/60Hz |
Outlet Amps | 3 |
Power Cord | 14 GA - 3 Wire, 15A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 6 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 55dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 320 (544) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 276-501 (649-851) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 903 |
Heat Rejected Vented (BTU/Hour) | 120 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49, EN 12469 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 460 lbs. / 209 kg. |
Net Weight lbs (kg) | 410 lbs. / 186 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Plate |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.99% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 3/4 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 59 (1499) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 87.375 (2219) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 93.375 (2371) |
Exterior Width Inches (mm) | 53 5/8 (1362) |
Nominal Width Feet (Meters) | 4 (1.2) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 45 7/8 (1165) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 46 3/8 (1178) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 2.9 |
Blower/Lights Amps [8-inch (203 mm) Access Opening] | 2.6 |
Electrical Configuration | 115V 60Hz |
Outlet Amps | 3 |
Power Cord | 12 gauge - 3 Wire, 20A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 14 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 56 dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 426 (724) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 363-588 (617-1000) |
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 359 (610) |
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 295-520 (502-884) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1140 |
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) | 1020 |
Heat Rejected Vented (BTU/Hour) | 157 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 530 lbs. / 240 kg. |
Net Weight lbs (kg) | 480 lbs. / 218 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Glass |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.995% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 3/4 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.995% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Access Opening(s) | 10-inch (254 mm) |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 59 (1499) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 31 7/16 (799) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 88 3/8 (2245) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 94 3/8 (2397) |
Exterior Width Inches (mm) | 53 5/8 (1362) |
Nominal Width Feet (Meters) | 4 ft. (1.2m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 45 7/8 (1165) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 46 3/8 (1178) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 2.9 |
Electrical Configuration | 230V 50/60Hz |
Outlet Amps | 3 |
Power Cord | 14 gauge - 3 Wire, 15A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 10 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 56 dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 426 (724) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 363-588 (617-1000) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1140 |
Heat Rejected Vented (BTU/Hour) | 157 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49, EN 12469 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 530 lbs. / 240 kg. |
Net Weight lbs (kg) | 480 lbs. / 218 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Plate |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.99% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 70 (1778) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 87.375 (2219) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 93.375 (2371) |
Exterior Width Inches (mm) | 65 5/8 (1669) |
Nominal Width Feet (Meters) | 5 ft. (1.5m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 57 7/8 (1470) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 46 3/8 (1178) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 4.5 |
Blower/Lights Amps [8-inch (203 mm) Access Opening] | 4.1 |
Electrical Configuration | 115V 60Hz |
Outlet Amps | 3 |
Power Cord | 12 gauge - 3 Wire, 20A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 16 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 58 dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 531 (902) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 451-676 (766-1149) |
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 445 (756) |
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 365-590 (621-1003) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1768 |
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) | 1611 |
Heat Rejected Vented (BTU/Hour) | 198 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 620 lbs. / 281 kg. |
Net Weight lbs (kg) | 570 lbs. / 258 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Glass |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.995% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.995% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 70 (1778) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 88 3/8 (2245) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 94 3/8 (2397) |
Exterior Width Inches (mm) | 65 5/8 (1669) |
Nominal Width Feet (Meters) | 5 ft. (1.5m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 57 7/8 (1470) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 46 3/8 (1178) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 2 |
Electrical Configuration | 230V 50/60Hz |
Outlet Amps | 3 |
Power Cord | 14 GA-3 Wire, 15 A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 11 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 58 |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 531 (902) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 451-676 (766-1149) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1768 |
Heat Rejected Vented (BTU/Hour) | 198 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49, EN 12469 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 620 lbs. / 281 kg. |
Net Weight lbs (kg) | 570 lbs. / 258 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Plate |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.99% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 84 (2134) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 87.375 (2219) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 93.375 (2371) |
Exterior Width Inches (mm) | 77 5/8 (1972) |
Nominal Width Feet (Meters) | 6 ft. (1.8m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 69 7/8 (1775) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 70 3/8 (1788) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 4.8 |
Blower/Lights Amps [8-inch (203 mm) Access Opening] | 4.5 |
Electrical Configuration | 115V 60Hz |
Outlet Amps | 3 |
Power Cord | 12 GA-3 Wire, 20A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 16 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 60 dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 634 (1077) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 538-763 (915-1297) |
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 532 (904) |
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 436-661 (741-1124) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1884 |
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) | 1768 |
Heat Rejected Vented (BTU/Hour) | 198 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 690 lbs. / 313 kg. |
Net Weight lbs (kg) | 640 lbs. / 290 kg. |
Cord Pass Through | Standard Right |
Outlet(s) | Standard, two backwall |
Plumbing | Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper |
Access Opening Glass Type | Safety Glass |
Armrest | Stainless Steel |
Cabinet Construction | All welded stainless steel 16/18 GA, Type 304 pressure tight design |
Decontamination Seal-able Window | Standard |
Diffuser | Metal, non-flammable |
Drain Valve Location | Front right |
Exhaust Filter | 11 1/2-inch (292 mm) 99.995% Efficiency @ 0.3 Micron |
Front & Control Panel Material | Painted Steel |
HEPA Filter Seal Type | HEPEX Zero Leak Neoprene Seal, Spring-loaded |
Lighting | LED |
Main Cabinet Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Motor(s) | 1 HP DC ECM |
Plenum Type | HEPEX |
Style | Bench top, optional base stand |
Supply Filter | 3-inch (76mm) 99.995% Efficiency @ 0.3 Micron |
Viewing Window Inches (mm) | 18 (457) open to fully closed |
Work Access Opening(s) | 10-inch (254 mm) |
Work Surface | One-Piece Dished Stainless Steel with Prop-Up Rod |
Work Zone Wrap | Single Piece Stainless Steel Side and Back Wall Construction |
Data Output Features | RS-232, USB |
Remote Alarm Terminals | Standard |
Est. Shipping Depth Inches (mm) | 43 (1092) |
Est. Shipping Height Inches (mm) | 71 (1803) |
Est. Shipping Width Inches (mm) | 84 (2134) |
Exterior Depth Inches (mm) | 33 7/8 (861) |
Exterior Depth with Armrest Removed Inches (mm) | 32 1/4 (819) |
Exterior Height (Minimum for Transport) Inches (mm) | 59 (1499) |
Exterior Height Inches (mm) | 60 7/8 (1546) |
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) | 88 3/8 (2245) |
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) | 94 3/8 (2397) |
Exterior Width Inches (mm) | 77 5/8 (1972) |
Nominal Width Feet (Meters) | 6 ft. (1.8m) |
Useable Work Tray Area Depth Inches (mm) | 20 7/8 (530) |
Useable Work Tray Area Width Inches (mm) | 69 7/8 (1775) |
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) | 26 (660) |
Work Area Height Inches (mm) | 28 1/2 (724) |
Work Area Width Inches (mm) | 70 3/8 (1788) |
Blower/Lights Amps [10-inch (254 mm) Access Opening] | 2.2 |
Electrical Configuration | 230V 50/60Hz |
Outlet Amps | 3 |
Power Cord | 14 GA-3 Wire, 15 A |
Power Cord Length Foot (Meter) | 12 (3.7) |
Rated Amps | 11 |
Light Intensity fc (Lux) | 90-120 (968-1291) |
Reach into workzone Inches (mm) | 6 (152) |
Sound Pressure Level per ISO 4871 | ≤ 60 dba |
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) | 634 (1077) |
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) | 538-763 (915-1297) |
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) | 1884 |
Heat Rejected Vented (BTU/Hour) | 198 |
Plant Duct Static Pressure Inches (mm) | 0.05-0.1"/1.27-2.54mm H2O |
Performance Standard | NSF/ANSI 49, EN 12469 |
Safety Certification | UL, UL-C |
Air Cleanliness (ISO 14644) | ISO Class 5 |
Downflow fpm (m/s) | 60 fpm (0.30 m/s) |
Fumigation: per NIH/NSF Procedures | True |
Inflow fpm (m/s) | 105 fpm (0.53 m/s) |
Protection | Product, Personnel, Environmental |
Airflow Monitor | Dual Airflow Sensor |
Control System | TouchLink |
Est. Shipping Weight lbs (kg) | 690 lbs. / 313 kg. |
Net Weight lbs (kg) | 640 lbs. / 290 kg. |
Review
Resources
Brochures
Manuals
Drawings
- NU-545-400 115V Biosafety Cabinet Cut Sheet
- 6-foot (1.8m) Biosafety Cabinet with Exhaust Canopy Cut Sheet
- NU-545-500E 230V Biosafety Cabinet Cut Sheet
- NU-545-600E 230V Biosafety Cabinet Cut Sheet
- NU-545-300E 230V Biosafety Cabinet Cut Sheet
- NU-545-400E 230V Biosafety Cabinet Cut Sheet
- NU-545-300 115V Biosafety Cabinet Cut Sheet
- NU-545-500 115V Biosafety Cabinet Cut Sheet
- NU-545-600 115V Biosafety Cabinet Cut Sheet
Specifications
BIM Revit
Request Your Quote for a NU-545
Thank you for completing the form.
We will be in touch. If you have any urgent questions, please reach out at 763-553-1270.