LabGard B1 NU-427 Class II Type B1 Biosafety Cabinet

The NU-427 preserves the Class II Type B1 cabinet’s traditional role of serving as an intermediate option between recirculating and total exhaust cabinets but gives this design a state-of-the-art control system.

Overview

The NU-427 preserves the Class II Type B1 cabinet’s traditional role of serving as an intermediate option between recirculating and total exhaust cabinets but with a control system fit for modern applications. By exhausting air from the entire length of the back half of the work zone through slots in the back wall, this design not only allows users to continue the recommended practice of working “clean to dirty” and deep inside the work zone but also leaves the dished work surface intact to retain spills. This exhaust is powered by an external facility blower that is interlocked with the cabinet’s own motor through a redundant system to provide two layers of protection that minimize the risk of blowing air out of the cabinet should the external blower fail: cabinet exhaust is monitored by not only a mechanical pressure switch but also a differential velocity pressure-flow grid. A touch screen displays the readings from both this flow grid and a downflow sensor in the work zone at all times for the cabinet operator. Standard models meet the requirements of the respected NSF/ANSI 49 standard for biosafety cabinets.

TouchLink Control System

The airflow dynamics in an externally exhausted type B2 cabinet need to work in constant harmony. The TouchLink control system displays inflow velocity, downflow velocity, and exhaust volume on a single, easy to read front screen. The digital control system interlocks the downflow blower with the external exhaust to shut down the motor if an exhaust failure occurs—a safeguard that is further backed up by a mechanical pressure switch.

TouchLink Electronic Control System for Class II, Type B2 Biosafety Cabinets

Digital Exhaust Flow Grid

The digital differential velocity pressure flow monitors exhaust to enable the TouchLink to derive not only inflow velocity but also the critical parameter of exhaust volume for display to the cabinet operator.

Digital exhaust flow grid for Class II, Type B Biosafety Cabinets

Downflow Air Sensor

The downflow sensor measures velocity inside the work zone and passes the data to the TouchLink display to assure users of product protection, a critical aspect in many pharmaceutical and biotechnology applications.

Biosafety Cabinet downflow thermistor airflow sensor.

Performance History

The TouchLink control panel showcases a week's worth of cabinet performance data on-screen, while also facilitating the download of a full month's operational history to a USB. This feature is essential for procedures requiring documented airflow conditions in compliance with Good Manufacturing Practice (GMP) standards.

Biosafety Cabinet historical performance graph on TouchLink Control System

Specifications

NU-427 Class II, Type B1 Biosafety Cabinet
Airflow Control Options Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels Optional Rear Closure Panels for Seismic Brackets
Cord Pass Through Standard on Right Side; Optional on Left Side
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) Optional 99.995% Efficient Metal-Frame Filter
Outlet(s) One, Left Front Faring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) One Coupling on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) One Valve on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light Optional
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting Fluorescent
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 1/2 HP DC ECM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 5-7/8 inch (149mm) 99.99% Efficiency on 0.3 Microns
Viewing Window Inches (mm) 18.5 (407) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel, Optional Prop-Up Rod
Work Zone Wrap N/A; Back Wall Inserted into Main Cabinet Wrap
Data Output Features USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 43 (1092)
Est. Shipping Height Inches (mm) 71 (1803)
Est. Shipping Width Inches (mm) 59 (1499)
Exterior Depth with Armrest Removed Inches (mm) 33 1/4 (845)
Exterior Height (Minimum for Transport) Inches (mm) 61 (1549)
Exterior Height Inches (mm) 61 3/8 (1559)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 87 7/8 (2232)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 93 7/8 (2385)
Exterior Width Inches (mm) 53 5/8 (1362)
Nominal Width Feet (Meters) 4 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 17 3/4 (451)
Useable Work Tray Area Width Inches (mm) 45 5/8 (1159)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 23 1/2 (597)
Work Area Height Inches (mm) 25 1/2 (648)
Work Area Width Inches (mm) 46 3/8 (1178)
Blower/Lights Amps 2.6
Electrical Configuration 115V60Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 10
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 282 (479)
Other Built-in Exhaust Option(s) Optional Charcoal Module, Optional 99.995% Efficient Metal-Frame Filter
Plant Duct Static Pressure Inches (mm) 0.7" w.g. / 18mm w.g
Performance Standard NSF-ANSI-49
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 70 fpm (0.36 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 543
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 600 (272)
Net Weight lbs (kg) 427 (249)
Airflow Control Options Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels Optional Rear Closure Panels for Seismic Brackets
Cord Pass Through Standard on Right Side; Optional on Left Side
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) Optional 99.995% Efficient Metal-Frame Filter
Outlet(s) One, Left Front Faring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) One Coupling on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) One Valve on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light Optional
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting Fluorescent
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 1/2 HP DC ECM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 5-7/8 inch (149mm) 99.99% Efficiency on 0.3 Microns
Viewing Window Inches (mm) 18.5 (407) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel, Optional Prop-Up Rod
Work Zone Wrap N/A; Back Wall Inserted into Main Cabinet Wrap
Data Output Features USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 43 (1092)
Est. Shipping Height Inches (mm) 71 (1803)
Est. Shipping Width Inches (mm) 59 (1499)
Exterior Depth with Armrest Removed Inches (mm) 33 1/4 (845)
Exterior Height (Minimum for Transport) Inches (mm) 61 (1549)
Exterior Height Inches (mm) 61 3/8 (1559)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 87 7/8 (2232)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 93 7/8 (2385)
Exterior Width Inches (mm) 53 5/8 (1362)
Nominal Width Feet (Meters) 4 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 17 3/4 (451)
Useable Work Tray Area Width Inches (mm) 45 5/8 (1159)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 23 1/2 (597)
Work Area Height Inches (mm) 25 1/2 (648)
Work Area Width Inches (mm) 46 3/8 (1178)
Blower/Lights Amps 1.3
Electrical Configuration 230V5060Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 10
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 282 (479)
Other Built-in Exhaust Option(s) Optional Charcoal Module, Optional 99.995% Efficient Metal-Frame Filter
Plant Duct Static Pressure Inches (mm) 0.7" w.g. / 18mm w.g
Performance Standard NSF-ANSI-49
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 70 fpm (0.36 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 774
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 600 (272)
Net Weight lbs (kg) 427 (249)
Airflow Control Options Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels Optional Rear Closure Panels for Seismic Brackets
Cord Pass Through Standard on Right Side; Optional on Left Side
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) Optional 99.995% Efficient Metal-Frame Filter
Outlet(s) One, Left Front Faring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) One Coupling on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) One Valve on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light Optional
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting Fluorescent
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 3/4 HP DCM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 5-7/8 inch (149mm)
Viewing Window Inches (mm) 18.5 (407) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel, Optional Prop-Up Rod
Work Zone Wrap N/A; Back Wall Inserted into Main Cabinet Wrap
Data Output Features USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 43 (1092)
Est. Shipping Height Inches (mm) 71 (1803)
Est. Shipping Width Inches (mm) 83 (2108)
Exterior Depth with Armrest Removed Inches (mm) 33 1/4 (845)
Exterior Height (Minimum for Transport) Inches (mm) 61 (1549)
Exterior Height Inches (mm) 61 3/8 (1559)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 87 7/8 (2232)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 93 7/8 (2385)
Exterior Width Inches (mm) 77 5/8 (1972)
Nominal Width Feet (Meters) 6 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 17 3/4 (451)
Useable Work Tray Area Width Inches (mm) 69 5/8 (1768)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 23 1/2 (597)
Work Area Height Inches (mm) 25 1/2 (648)
Work Area Width Inches (mm) 70 3/8 (1788)
Blower/Lights Amps 4.5
Electrical Configuration 115V60Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 12
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 474 (805)
Other Built-in Exhaust Option(s) Optional Charcoal Module, Optional 99.995% Efficient Metal-Frame Filter
Plant Duct Static Pressure Inches (mm) 0.9" w.g. / 23mm w.g.
Performance Standard NSF-ANSI-49
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 70 fpm (0.36 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 774
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 790 (358)
Net Weight lbs (kg) 740 (336)
Airflow Control Options Cabinet Off Override, Delay Off Exhaust Airflow, Exhaust Airflow Loss Override
Blower Contact Wires N/A
Closure Panels Optional Rear Closure Panels for Seismic Brackets
Cord Pass Through Standard on Right Side; Optional on Left Side
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks Optional
Motorized Base Stand Optional as Separate Accessory
Motorized Window N/A
Other Built-in Supply Filter Option(s) Optional 99.995% Efficient Metal-Frame Filter
Outlet(s) One, Left Front Faring
Plumbing Optional 3/8" Soft Copper, 3/8" Black Pipe (Gas Only), or 1/2" Rigid Copper
Remote Service Valves Maximum of Three Each Side Wall Optional
Seismic Brackets or Studs Optional Rear or Floor Mount Seismic Brackets
Service Couplings (3/8-inch NPT) One Coupling on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Service Valves (3/8-inch NPT) One Valve on Right Side Wall is Standard, Maximum of Three Each Side Wall Optional
Side Panels Required with Plumbing out Bottom, Top or Back
Smooth Interior N/A
Ultraviolet (UV) Light Optional
Window Edge Protector Optional
Access Opening Glass Type Tempered Glass
Cabinet Construction All welded stainless steel 16/18 GA, Type 304 pressure tight design
Casters N/A
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 11 1/2-inch (292 mm) 99.99% Efficiency on 0.3 Microns
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting Fluorescent
Main Cabinet Wrap Single Piece Stainless Steel Side and Back Wall Construction
Motor(s) 3/4 HP DCM
Plenum Type HEPEX
Pre-filter N/A
Style Bench top, optional base stand
Supply Filter 5-7/8 inch (149mm)
Viewing Window Inches (mm) 18.5 (407) open to fully closed
Work Access Opening(s) 8-inch (203mm)
Work Surface One-Piece Dished Stainless Steel, Optional Prop-Up Rod
Work Zone Wrap N/A; Back Wall Inserted into Main Cabinet Wrap
Data Output Features USB
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 43 (1092)
Est. Shipping Height Inches (mm) 71 (1803)
Est. Shipping Width Inches (mm) 83 (2108)
Exterior Depth with Armrest Removed Inches (mm) 33 1/4 (845)
Exterior Height (Minimum for Transport) Inches (mm) 61 (1549)
Exterior Height Inches (mm) 61 3/8 (1559)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 87 7/8 (2232)
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 93 7/8 (2385)
Exterior Width Inches (mm) 77 5/8 (1972)
Nominal Width Feet (Meters) 6 ft. (1.2m)
Useable Work Tray Area Depth Inches (mm) 17 3/4 (451)
Useable Work Tray Area Width Inches (mm) 69 5/8 (1768)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) 23 1/2 (597)
Work Area Height Inches (mm) 25 1/2 (648)
Work Area Width Inches (mm) 70 3/8 (1788)
Blower/Lights Amps 2.2
Electrical Configuration 230V5060Hz
Junction Box Optional
Outlet Amps 3
Power Cord 14 GA - 3 Wire, 15A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 11
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
Bag In Bag Out Optional
Direct Connection Exhaust Requirement Concurrent Balance Value, CFM(CMH) 474 (805)
Other Built-in Exhaust Option(s) Optional Charcoal Module, Optional 99.995% Efficient Metal-Frame Filter
Plant Duct Static Pressure Inches (mm) 0.9" w.g. / 23mm w.g.
Performance Standard NSF-ANSI-49
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 70 fpm (0.36 m/s)
Fumigation: per NIH/NSF Procedures True
Heat Rejected (BTU/Hour) 774
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor AirflowSensorPressureGrid
Control System TouchLink
Est. Shipping Weight lbs (kg) 790 (358)
Net Weight lbs (kg) 740 (336)

Review

Options

  • Coved Corner Tempered Glass 8 Inch Sliding Window
  • Coved Corner Tempered Glass 8" Sliding Window
  • Tempered Glass 8 Inch Sliding Window with Standard Divider
  • Cabinet Off Override
  • Delay Off Exhaust Airflow
  • Exhaust Airflow Loss Override
  • Plastic Armrest
  • Stainless Steel Armrest
  • IV Bar With 6 Hooks
  • Left Sidewall Cord Pass-Through
  • Right Sidewall NSF Cord Pass-Through
  • Stainless Steel Window Edge Protector
  • Worktray Prop-Up
  • 99.99% Wood Frame Exhaust Filter
  • 99.995% Metal Frame Exhaust Filter
  • Charcoal Exhaust Filter Module
  • Bag-In/Bag-Out Exhaust HEPA Filter
  • Standard Configuration Without Bag-In/Bag-Out Exhaust Filter
  • Without Standard Bag-In/Bag-Out Exhaust Filter
  • 115V Flush Mount Front Decorative Panel
  • 115V 60Hz Standard Outlet
  • 115V Ground Fault Interrupter
  • 2 Duplex Outlets With GFI Option
  • Add 2 Faring Outlets
  • 1/2" RIGID COPPER
  • 3/8" Black Pipe
  • 3/8" Soft Copper
  • 115V 60Hz 15Amp Power Input
  • Electrical Junction Box
  • Bottom Left Remote Control Valve
  • Bottom Right Remote Control Valve
  • Middle Left Remote Control Valve
  • Middle Right Remote Control Valve
  • Top Left Remote Control Valve
  • Top Right Remote Control Valve
  • 28 Inch Seismic Floor Brackets
  • Rear Mount Seismic Bracket For Adjustable Basestand
  • Rear Mount Seismic Bracket For Fixed Basestand
  • Rear Mount Seismic Brackets For Adjustable Basestand
  • Rear Mount Seismic Brackets For Fixed Height Basestand
  • Side Mount Seismic Brackets For Adjustable Basestand
  • Side Mount Seismic Brackets For Telescoping Basestand
  • Bottom Left Plugged Coupling
  • Bottom Right Plugged Coupling
  • Middle Left Plugged Coupling
  • Middle Right Plugged Coupling
  • Top Left Plugged Coupling
  • Top Right Plugged Coupling
  • Bottom Left Service Valve
  • Bottom Right Service Valve
  • Middle Left Service Valve
  • Middle Right Service Valve
  • Top Left Service Valve
  • Top Right Service Valve
  • Side Panels With Plumbing Out Back
  • Side Panels With Plumbing Out Bottom
  • Side Panels With Plumbing Out Top
  • Side Panels Without Plumbing
  • Metal Frame Hepex Bag
  • 115V Ultraviolet Light
  • Without Ultraviolet (UV) Light

Request Your Quote for a NU-427

Please fill out the form below for a lab equipment quote and we'll get to work on getting you what you need. Need to talk with someone right away? Give us a call at 763-553-1270.

Thank you for completing the form.

We will be in touch. If you have any urgent questions, please reach out at 763-553-1270.