LabGard LP NU-640 Class II Type A2 Biosafety Cabinet

The console style NU-640 biological safety cabinet features the tall work access opening needed to work with large cages but in a low-profile configuration designed to fit in vivariums with low ceiling heights.

Overview

The console style NU-640 Class II, Type A2 animal handling biosafety cabinet features the tall work access opening needed to work with large cages in a low-profile configuration designed to fit in vivariums with low ceiling heights. This versatile design’s motorized height adjustment eases installation into otherwise inaccessible vivariums and adaptation to individual cabinet users’ ergonomic needs. A built-in footrest enhances comfort for standing operators. Built for a broad range of cage changing applications, the hinged window with a 14-inch (356mm) access opening allows a technician to introduce rat cages into the work zone with ample arm room to reach the interior area of the cage. The flat work surface can be easily raised and rested on support rods to access and maintain the pre-filter located underneath. The steel construction of the cabinet forms a core structure that will endure years of use. Simultaneously, an Ultra-high efficiency motor located at the bottom of the console design ensures enough reserve power to compensate for filter loading. The Aeromax™ control system visualizes plenum pressure and cabinet performance through a series of LED lights, raising an alarm if necessary.

AeroMax Control System

The AeroMax control system visualizes airflow plenum pressure through a series of multicolored lights and activates both an audible and visual alarm whenever cabinet performance departs from normal parameters.

AeroMax Electronic Control System for Lab Equipment

Ergonomic Efficiency

The NU-640, with its innovative console model design, stands out as the quintessential solution for vivariums challenged by low ceiling heights. Its low-profile structure is specifically crafted to fit into spaces where traditional cabinets wouldn't, offering unmatched adaptability. However, it's not just about fitting in; the NU-640 prioritizes operator comfort and safety. Its unique design includes an adjustable work surface height feature, ensuring that technicians of any stature can find their perfect working position. This focus on ergonomics is crucial in reducing workplace injuries associated with repetitive tasks, a common concern in laboratory environments. With the NU-640, vivariums no longer have to compromise on space, functionality, or staff well-being.

Lab animal tech cage changing in a LabGard NU-640 Class II, Type A2 Animal Handling Biosafety Cabinet

Industry Leading Work Access Opening

The NU-640 revolutionizes vivarium operations with its unprecedentedly large access opening, specifically designed to accommodate the effortless entry and removal of even the largest rat cages. This expansive opening removes the everyday struggle of maneuvering bulky items in and out of a confined space, making the process seamless. The cabinet's enhanced 14-inch (356mm) work access opening height is a game-changer, significantly expanding the usable work zone. This innovative feature provides lab animal technicians with the invaluable flexibility to move their arms and hands freely above the cages, substantially simplifying cage transfers. With the NU-640, routine tasks are no longer cumbersome, thanks to the thoughtfully designed access dimensions that transform operational efficiency and ease.

Lab animal technician reaching hand into mouse cage in a biosafety cabinet.

Pre-filter Extends Life and Lowers Cost

The durability of primary supply and exhaust HEPA filters is paramount. Here, the unique challenge is the presence of animal hair and dander, which can swiftly overwhelm and reduce the efficacy of standard filters. However, this concern is effectively addressed with a specialized, reusable pre-filter positioned strategically beneath the work surface. This pre-filter is engineered to capture larger particulates, safeguarding the primary HEPA filters and allowing them to function more efficiently against finer contaminants. Not only does this extend the life of the HEPA filters, but the reusability of the pre-filter also curtails the need for frequent and potentially expensive replacements. Additionally, the cabinet's innovative design features a prop-up work surface, granting easy access for prompt and hassle-free cleaning and maintenance of the pre-filter. This thoughtful integration of features underscores a commitment to operational longevity and cost-effectiveness.

LabGard NU-640 Animal Handling Biosafety Cabinet pre-filter under the work surface.

Specifications

NU-640 Animal Handling Class II, Type A2 Biosafety Cabinet
Blower Contact Wires N/A
Closure Panels N/A
Cord Pass Through N/A
Double Exhaust HEPA Filter N/A
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks N/A
Magnifying Glass in Window N/A
Motorized Base Stand Included
Motorized Window N/A
Other Built-in Supply Filter Option(s) N/A
Outlet(s) Standard, two backwall
Plumbing N/A
Remote Service Valves N/A
Seismic Brackets or Studs N/A
Service Couplings (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Service Valves (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Side Panels N/A
Smooth Interior N/A
Ultraviolet (UV) Light N/A
Window Edge Protector Optional
Access Opening Glass Type Safety Glass
Armrest Optional, Stainless Steel
Cabinet Construction All welded stainless steel 16/18 gauge, Type 304 pressure tight design
Casters 5" (13mm) Casters Standard
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 6-inch (152 mm) 99.99% Efficiency @ 0.3 Micron
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting LED
Main Cabinet Wrap N/A-Multi-Piece Steel Construction
Motor(s) 3/4 HP DC ECM
Plenum Type HEPEX
Pre-filter Underneath Work Surface
Style Console
Supply Filter 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron
Viewing Window Inches (mm) 23.875 (607) Safety Glass/Hinged
Work Access Opening(s) 14-inches (356 mm)
Work Surface One-Piece Flat Stainless Steel with Prop-Up Rod
Work Zone Wrap Single Piece Stainless Steel Side and Back Wall Construction
Data Output Features RS-232
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 39 (991)
Est. Shipping Height Inches (mm) 87 (2210)
Est. Shipping Width Inches (mm) 55 (1397)
Exterior Depth Inches (mm) 32 7/8 (835)
Exterior Depth with Armrest Removed Inches (mm) 33 (838)
Exterior Height (Minimum for Transport) Inches (mm) 78.25 (1987)
Exterior Height Inches (mm) 78 1/4-86 1/4 (1987-2190)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 78 1/4 (1987) with 35-in (890mm) Work Surface Height
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 86 1/4 (2190) with 43-in (1093) Work Surface Height
Exterior Width Inches (mm) 56.375 (1433) Including push/pull bars
Nominal Width Feet (Meters) 4 (1.2)
Useable Work Tray Area Depth Inches (mm) 19 (482)
Useable Work Tray Area Width Inches (mm) 46.2 (1173)
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 12-inches (305 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 14-inches (356 mm) Window Height Inches (mm) 22.75 (579)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) N/A
Work Area Height Inches (mm) 24.625 (626)
Work Area Width Inches (mm) 46.5 (1181)
Blower/Lights Amps [10-inch (254 mm) Access Opening]
Blower/Lights Amps [12-inch (305 mm) Access Opening]
Blower/Lights Amps [14-inch (356mm) Access Opening] 4.8
Blower/Lights Amps [8-inch (203 mm) Access Opening]
Electrical Configuration 115V60Hz
Junction Box N/A
Outlet Amps 3
Power Cord 12 gauge - 3 Wire, 20A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 14
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
14-inch (356mm) Opening Exhaust Requirement for Fixed Canopy (NU-926) CFM(CMH) 609 (1035)
14-inch (356mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM(CMH) 575-725 (977-1232)
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
Bag In Bag Out N/A
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 12-inch (305 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 14-inch (356 mm) Access Opening (BTU/Hour) 1883
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) N/A
Heat Rejected Vented (BTU/Hour) 198
Other Built-in Exhaust Option(s) N/A
Plant Duct Static Pressure Inches (mm) 0.05-0.1"/1.27-2.54mm H2O
Australian Test Method N/A
Performance Standard
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 60 fpm (0.30 m/s)
Fumigation: per NIH/NSF Procedures True
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor Pressure
Control System AeroMax
Est. Shipping Weight lbs (kg) 610 lbs. / 277 kg.
Net Weight lbs (kg) 560 lbs. / 254 kg.
Blower Contact Wires N/A
Closure Panels N/A
Cord Pass Through N/A
Double Exhaust HEPA Filter N/A
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks N/A
Magnifying Glass in Window N/A
Motorized Base Stand Included
Motorized Window N/A
Other Built-in Supply Filter Option(s) N/A
Outlet(s) Standard, two backwall
Plumbing N/A
Remote Service Valves N/A
Seismic Brackets or Studs N/A
Service Couplings (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Service Valves (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Side Panels N/A
Smooth Interior N/A
Ultraviolet (UV) Light N/A
Window Edge Protector Optional
Access Opening Glass Type Safety Glass
Armrest Optional, Stainless Steel
Cabinet Construction All welded stainless steel 16/18 gauge, Type 304 pressure tight design
Casters 5" (13mm) Casters Standard
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 6-inch (152 mm) 99.99% Efficiency @ 0.3 Micron
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting LED
Main Cabinet Wrap N/A-Multi-Piece Steel Construction
Motor(s) 1 HP DC ECM
Plenum Type HEPEX
Pre-filter Underneath Work Surface
Style Console
Supply Filter 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron
Viewing Window Inches (mm) 23.875 (607) Safety Glass/Hinged
Work Access Opening(s) 14-inch (356 mm)
Work Surface One-Piece Flat Stainless Steel with Prop-Up Rod
Work Zone Wrap Single Piece Stainless Steel Side and Back Wall Construction
Data Output Features RS-232
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 39 (991)
Est. Shipping Height Inches (mm) 87 (2210)
Est. Shipping Width Inches (mm) 73 (1854)
Exterior Depth Inches (mm) 32 7/8 (835)
Exterior Depth with Armrest Removed Inches (mm) 33 (838)
Exterior Height (Minimum for Transport) Inches (mm) 78 1/4 (1987)
Exterior Height Inches (mm) 78 1/4-86 1/4 (1987-2190)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 78 1/4 (1987) with 35-in (890mm) Work Surface Height
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 86 1/4 (2190) with 43-in (1093) Work Surface Height
Exterior Width Inches (mm) 68 3/8 (1737) including push/pull bars
Nominal Width Feet (Meters) 5 ft. (1.5m)
Useable Work Tray Area Depth Inches (mm) 19 (482)
Useable Work Tray Area Width Inches (mm) 58.2 (1478)
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 12-inches (305 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 14-inches (356 mm) Window Height Inches (mm) 22 3/4 (579)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) N/A
Work Area Height Inches (mm) 24 5/8 (626)
Work Area Width Inches (mm) 58 1/2 (1486)
Blower/Lights Amps [10-inch (254 mm) Access Opening]
Blower/Lights Amps [12-inch (305 mm) Access Opening]
Blower/Lights Amps [14-inch (356mm) Access Opening] 5.6
Blower/Lights Amps [8-inch (203 mm) Access Opening]
Electrical Configuration 115V60Hz
Junction Box N/A
Outlet Amps 3
Power Cord 12 ft. 12 GA-3 Wire, 20A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 16
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
14-inch (356mm) Opening Exhaust Requirement for Fixed Canopy (NU-926) CFM(CMH) 731 (1242)
14-inch (356mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM(CMH) 699-829 (1188-1408)
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
Bag In Bag Out N/A
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 12-inch (305 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 14-inch (356 mm) Access Opening (BTU/Hour) 2197
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) N/A
Heat Rejected Vented (BTU/Hour) 198
Other Built-in Exhaust Option(s) N/A
Plant Duct Static Pressure Inches (mm) 0.05-0.1"/1.27-2.54mm H2O
Australian Test Method N/A
Performance Standard
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 60 fpm (0.30 m/s)
Fumigation: per NIH/NSF Procedures True
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor Pressure
Control System AeroMax
Est. Shipping Weight lbs (kg) 700 (318)
Net Weight lbs (kg) 650 (295)
Blower Contact Wires N/A
Closure Panels N/A
Cord Pass Through N/A
Double Exhaust HEPA Filter N/A
Gas Supply Interlock with Blower N/A
Glass Sidewalls N/A
IV Bar with 6 Hooks N/A
Magnifying Glass in Window N/A
Motorized Base Stand Included
Motorized Window N/A
Other Built-in Supply Filter Option(s) N/A
Outlet(s) Standard, two backwall
Plumbing N/A
Remote Service Valves N/A
Seismic Brackets or Studs N/A
Service Couplings (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Service Valves (3/8-inch NPT) None, Maximum of Two Each Side Wall Optional
Side Panels N/A
Smooth Interior N/A
Ultraviolet (UV) Light N/A
Window Edge Protector Optional
Access Opening Glass Type Safety Glass
Armrest Optional, Stainless Steel
Cabinet Construction All welded stainless steel 16/18 gauge, Type 304 pressure tight design
Casters 5" (13mm) Casters Standard
Decontamination Seal-able Window N/A
Diffuser Metal, non-flammable
Drain Valve Location Front right
Exhaust Filter 6-inch (152 mm) 99.99% Efficiency @ 0.3 Micron
Front & Control Panel Material Painted Steel
HEPA Filter Seal Type HEPEX Zero Leak Neoprene Seal, Spring-loaded
Lighting LED
Main Cabinet Wrap N/A-Multi-Piece Steel Construction
Motor(s) 1 HP DC ECM
Plenum Type HEPEX
Pre-filter Underneath Work Surface
Style Console
Supply Filter 3-inch (76mm) 99.99% Efficiency @ 0.3 Micron
Viewing Window Inches (mm) 23.875 (607) Safety Glass/Hinged
Work Access Opening(s) 14-inch (356)
Work Surface One-Piece Flat Stainless Steel with Prop-Up Rod
Work Zone Wrap Single Piece Stainless Steel Side and Back Wall Construction
Data Output Features RS-232
Remote Alarm Terminals Standard
Est. Shipping Depth Inches (mm) 39 (991)
Est. Shipping Height Inches (mm) 87 (2210)
Est. Shipping Width Inches (mm) 81 (2057)
Exterior Depth Inches (mm) 32 7/8 (835)
Exterior Depth with Armrest Removed Inches (mm) 33 (838)
Exterior Height (Minimum for Transport) Inches (mm) 78 1/4 (1987)
Exterior Height Inches (mm) 78 1/4-86 1/4 (1987-2190)
Exterior Height with 30-inch (762 mm) Work Surface Height inches (mm) 78 1/4 (1987) with 35-in (890mm) Work Surface Height
Exterior Height with 36-inch (915 mm) Work Surface Height inches (mm) 86 1/4 (2190) with 43-in (1093) Work Surface Height
Exterior Width Inches (mm) 80 3/8 (2042)
Nominal Width Feet (Meters) 6 ft. (1.8m)
Useable Work Tray Area Depth Inches (mm) 19 (482)
Useable Work Tray Area Width Inches (mm) 70 (1783)
Work Area Depth Measured at 10-inches (254 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 12-inches (305 mm) Window Height Inches (mm) N/A
Work Area Depth Measured at 14-inches (356 mm) Window Height Inches (mm) 22 3/4 (579)
Work Area Depth Measured at 8-inches (203 mm) Window Height Inches (mm) N/A
Work Area Height Inches (mm) 24 5/8 (626)
Work Area Width Inches (mm) 70 1/2 (1791)
Blower/Lights Amps [10-inch (254 mm) Access Opening]
Blower/Lights Amps [12-inch (305 mm) Access Opening]
Blower/Lights Amps [14-inch (356mm) Access Opening] 6.0
Blower/Lights Amps [8-inch (203 mm) Access Opening]
Electrical Configuration 115V60Hz
Junction Box N/A
Outlet Amps 3
Power Cord 12 ft. 12 GA-3 Wire, 20A
Power Cord Length Foot (Meter) 12 (3.7)
Rated Amps 16
Light Intensity fc (Lux) 90-120 (968-1291)
Reach into workzone Inches (mm) 6 (152)
Sound Pressure Level per ISO 4871 Inquire
10-inch (254 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
10-inch (254 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
12-inch (305 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
14-inch (356mm) Opening Exhaust Requirement for Fixed Canopy (NU-926) CFM(CMH) 888 (1509)
14-inch (356mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM(CMH) 880-970 (1495-1648)
8-inch (203 mm) Opening Exhaust Requirement for Fixed Canopy (NU-907/NU-916) CFM (CMH) N/A
8-inch (203 mm) Opening Exhaust Requirement for Variable Flow Canopy (NU-911) CFM (CMH) N/A
Bag In Bag Out N/A
Heat Rejected 10-inch (254 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 12-inch (305 mm) Access Opening (BTU/Hour) N/A
Heat Rejected 14-inch (356 mm) Access Opening (BTU/Hour) 2354
Heat Rejected 8-inch (203 mm) Access Opening (BTU/Hour) N/A
Heat Rejected Vented (BTU/Hour) 2354
Other Built-in Exhaust Option(s) N/A
Plant Duct Static Pressure Inches (mm) 0.05-0.1"/1.27-2.54mm H2O
Australian Test Method N/A
Performance Standard
Safety Certification
Air Cleanliness (ISO 14644) ISO-Class-5
Downflow fpm (m/s) 60 fpm (0.30 m/s)
Fumigation: per NIH/NSF Procedures True
Inflow fpm (m/s) 105 fpm (0.53 m/s)
Protection Product, Personnel, Environmental
Airflow Monitor Pressure
Control System AeroMax
Est. Shipping Weight lbs (kg) 790 (358)
Net Weight lbs (kg) 740 (336)

Review

Request Your Quote for a NU-640

Please fill out the form below for a lab equipment quote and we'll get to work on getting you what you need. Need to talk with someone right away? Give us a call at 763-553-1270.

Thank you for completing the form.