In-VitroCell NU-5820 Direct HeatCO2 Incubator
- Overview
- Specifications
- Resources
Overview
The In-VitroCell CO₂ Incubator model NU-5820 features humidity control through a relative humidity (RH) sensor and a water reservoir to minimize cell desiccation. The RH sensor monitors humidity levels in the growth chamber. When humidity is needed to maintain setpoint, vapor is injected into the chamber from the water reservoir. In addition, model NU-5820 comes standard with a 200-liter capacity, dual decontamination cycles, and a dual wave single source infrared (IR) sensor. In-VitroCell NU-5820 creates optimal cell growth conditions and is the perfect solution for cell lines prone to desiccation.
Optimal Growth Conditions with Effortless Relative Humidity Control
Achieve optimal growth conditions with our advanced product feature: Relative humidity levels of up to 90% can be effortlessly attained in the growth chamber. This is made possible by utilizing a 3-liter reservoir located externally to the chamber. The chamber air is intelligently directed through an evaporate box, where it comes into contact with water vapor-saturated air. This process ensures that humidity levels are adjusted accordingly, based on the feedback provided by the RH sensor. Experience the perfect balance of science and user-friendliness with our innovative solution.
Simplified Precision: NuTouch Control System
The NuTouch control system offers an intuitive and efficient user experience, designed with your ease and convenience in mind. With this advanced interface, you can effortlessly adjust set points, access chamber history, initiate a decontamination cycle, or carry out calibration procedures. What's more, all NuAire CO2 incubators feature the NuTouch control system, ensuring a consistent and seamless experience across multiple units in your laboratory. Your journey with our incubators just got smarter and more user-friendly.
Cutting-Edge Dual Wave Infrared (IR) Sensor
Our state-of-the-art CO2 control system utilizes a sophisticated dual-wave infrared (IR) sensor that's specifically designed to maintain the perfect pH levels within the growth chamber. This sensor operates at wavelengths that are selectively absorbed solely by CO2, ensuring that measurements remain unaffected by other elements like water vapor. Its advanced design guarantees a stable output, greatly reducing the need for frequent calibrations. With this technology, your experiments can thrive in the most precise and controlled environment.
Specifications
| Inner Doors | One |
| Exterior Depth Inches (mm) | 27.188 (691) |
| Exterior Height Inches (mm) | 39.688 (1008) |
| Exterior Width Inches (mm) | 26.750 (680) |
| Footprint Depth inches (mm) | 20.75 (527) |
| Footprint Width inches (mm) | 23.75 (603) |
| Interior Depth Inches (mm) | 20.053 (509) |
| Interior Height Inches (mm) | 28.500 (724) |
| Interior Volume Cubic Feet (Liters) | 7.06 (200) |
| Interior Width Inches (mm) | 21.257 (540) |
| Electrical Configuration | 115V 60Hz |
| Power Cord Length Foot (Meter) | 8 (2.5) |
| Safety Certification | UL, UL-C |
| Decontamination Cycle | 95°C Humidified Decontamination, 145°C Dry Decontamination Cycle |
| RH Accuracy | ± 3% |
| RH Control Logic | Fixed algorithm/manual environmental adaptable |
| RH Control Range | Ambient dependent to 90% |
| RH Display Resolution | 0.1% |
| RH Recovery | 90% +5%/-3% within 25 minutes on average |
| RH Sensor Type | Precision capacitance bulk-polymer |
| Control System | NuTouch |
| Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable] |
| Temperature Accuracy °C | ± 0.1% |
| Temperature Control | Direct Heat |
| Temperature Control Range | 5°C above ambient (to a 30°C max ambient) to 55°C |
| Temperature Display Resolution | 0.1°C |
| Temperature Recovery | 0.12°C/min. on Average |
| Temperature Sensor Type | Precision Integrated Circuit |
| Temperature Set-Point Range | 5°C to 55°C (37.0°C Default) |
| Temperature Uniformity °C | ± 0.35°C @ 37°C |
| Net Weight lbs (kg) | 235 lbs. / 106 kg. |
| Inner Doors | One |
| Exterior Depth Inches (mm) | 27.188 (691) |
| Exterior Height Inches (mm) | 39.688 (1008) |
| Exterior Width Inches (mm) | 26.750 (680) |
| Footprint Depth inches (mm) | 20.75 (527) |
| Footprint Width inches (mm) | 23.75 (603) |
| Interior Depth Inches (mm) | 20.053 (509) |
| Interior Height Inches (mm) | 28.500 (724) |
| Interior Volume Cubic Feet (Liters) | 7.06 (200) |
| Interior Width Inches (mm) | 21.257 (540) |
| Electrical Configuration | 230V 50/60Hz |
| Power Cord Length Foot (Meter) | 8 (2.5) |
| Safety Certification | UL, UL-C |
| Decontamination Cycle | 95°C Humidified Decontamination, 145°C Dry Decontamination Cycle |
| RH Accuracy | ± 3% |
| RH Control Logic | Fixed algorithm/manual environmental adaptable |
| RH Control Range | Ambient dependent to 90% |
| RH Display Resolution | 0.1% |
| RH Recovery | 90% +5%/-3% within 25 minutes on average |
| RH Sensor Type | Precision capacitance bulk-polymer |
| Control System | NuTouch |
| Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable] |
| Temperature Accuracy °C | ± 0.1% |
| Temperature Control | Direct Heat |
| Temperature Control Range | 5°C above ambient (to a 30°C max ambient) to 55°C |
| Temperature Display Resolution | 0.1°C |
| Temperature Recovery | 0.12°C/min. on Average |
| Temperature Sensor Type | Precision Integrated Circuit |
| Temperature Set-Point Range | 5°C to 55°C (37.0°C Default) |
| Temperature Uniformity °C | ± 0.35°C @ 37°C |
| Net Weight lbs (kg) | 235 lbs. / 106 kg. |